ORIGINAL ARTICLE |
|
Year : 2009 | Volume
: 3
| Issue : 2 | Page : 66-69 |
|
Quality evaluation of four hemoglobin screening methods in a blood donor setting along with their comparative cost analysis in an Indian scenario
Rashmi Tondon, Anupam Verma, Prashant Pandey, Rajendra Chaudhary
Department of Transfusion Medicine, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow- 226014, UP, India
Correspondence Address:
Anupam Verma Department of Transfusion Medicine, Sanjay Gandhi Postgraduate Institute of Medical Sciences, Lucknow- 226014, UP India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0973-6247.53874
|
|
Background: Despite the wide range of methods available for measurement of hemoglobin, no single technique has emerged as the most appropriate and ideal for a blood donation setup. Materials and Methods: A prospective study utilizing 1014 blood samples was carried out in a blood donation setting for quality evaluation of four methods of hemoglobin estimation along with cost analysis: Hematology cell analyzer (reference), HCS, CuSO4 method and HemoCue. Results: Mean value of HemoCue (mean ± SD = 14.7 ± 1.49 g/dl) was higher by 0.24 compared to reference (mean ± SD = 13.8 ± 1.52 g/dl) but not statistically significant ( P > 0.05). HemoCue proved to be the best technique (sensitivity 99.4% and specificity 84.4%) whereas HCS was most subjective with 25.2% incorrect estimations. CuSO4 proved to be good with 7.9% false results. Comparative cost analysis of each method was calculated to be 35 INR/test for HemoCue, 0.76 INR /test for HCS and 0.06-0.08 INR /test for CuSO4. Conclusion: CuSO4 method gives accurate results, if strict quality control is applied. HemoCue is too expensive to be used as a primary screening method in an economically restricted country like India. |
|
|
|
[FULL TEXT] [PDF]* |
|
 |
|