Evaluation of the pneumatic tube system for transportation of packed red cell units


Department of Transfusion Medicine, Tata Medical Center, Kolkata, West Bengal, India

Correspondence Address:
Dr. Sabita Basu
Department of Transfusion Medicine, Tata Medical Center, 14 MAR (EW), New Town, Rajarhat, Kolkata - 700 156, West Bengal
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0973-6247.154254

Rights and Permissions

Background: Pneumatic tube system (PTS) is commonly used in hospital settings to transport blood samples to diagnostic laboratories. At our blood center, we receive blood requisitions via the PTS, but units are carried to the ward by human courier. Recently we considered using the PTS for transporting blood units. Since, there are reports of hemolysis in blood samples sent through the PTS, we evaluated this system for transporting red cell units. Aims: The aim was to assess the effect of PTS transport on the quality of packed red cell units. Materials and Methods: A total of 50 red blood cells units (RBC), (25 non-irradiated and 25 irradiated) were subjected to transportation through the PTS. The control arm in the study was age-matched RBC units not subjected to PTS transport. Each RBC unit was evaluated for hemoglobin (Hb), lactate dehydrogenase, potassium and plasma hemoglobin (Hb). The paired t-test was used to compare these parameters, and the P value was calculated. Results and Conclusion: The percentage of hemolysis after transportation through PTS was below the recommended guidelines. Delivery of the blood unit to the wrong station, bags lying unattended at the destination were few of the problems that had to be addressed. To conclude, though the PTS is a safe means of transporting blood products with reduction in the turn-around-time, it must be validated before use.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed4637    
    Printed119    
    Emailed1    
    PDF Downloaded65    
    Comments [Add]    
    Cited by others 2    

Recommend this journal

 

Association Contact us | Sitemap | Advertise | What's New | Copyright and Disclaimer | Privacy Notice


2006 - Asian Journal of Transfusion Science | Published by Wolters Kluwer - Medknow
Online since 10th November, 2006